CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8500A/AEOPT1/2/5/8 as used with the TS 3846/ASM608 Inertial Measuring Unit Test Set
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
4.1.9
1 V
"
100 V
"
1000 V
TI 0.1 V Range
1 V LSD
1 V LSD
4.2
Alternating–Voltage Measurements (01)
TI Range
4.2.2
1 V
"
10 V
"
100 V
"
1000 V
"
10 V
"
10 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00000
10 V x .1
1.00000
10 V x .2
2.00000
10 V x .3
3.00000
10 V x .4
4.00000
10 V x .5
5.00000
10 V x .6
6.00000
10 V x .7
7.00000
10 V x .8
8.00000
10 V x .9
9.00000
10 V x 1
10.00000
10 V x 1
–10.00000
1 V x 1
1.00000
100 V
100.0000
100 V
100.0000
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
400 Hz–100 V
100.000
400 Hz–100 V
100.000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0000–1 to 0.0000–1
0.9991–0
to 1.0009–0
1.9991–0
to 2.0009–0
2.9991–0
to 3.0009–0
3.9991–0
to 4.0009–0
4.9991–0
to 5.0009–0
5.9991–0
to 6.0009–0
6.9991–0
to 7.0009–0
7.9991–0
to 8.0009–0
8.9991–0
to 9.0009–0
9.9991–0
to 10.0009–0
–9.9991–0 to 10.0009–0
0.9999–1
to 1.0000–9
99.991–0
to 100.009–0
99.910–0
to 100.090–0
( V)
501
to 503
497
to 499
(V)
0.99850
to 1.00150
9.9880
to 10.0120
99.880
to 100.120
98.200
to 101.800
9.9880
to 10.0120
9.9005
to 10.1005
Page 1 of 2