CALIBRATION CHECKLIST
TEST INST (S)
Cubic 242 Instrument System
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
4.1.9
100 V
"
1000 V
Affix a special calibration label stating:
"The ratio measurement function has not been calibrated."
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
100
100.00
100
100.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.001
to
0.001
0.999
to
1.001
1.999
to
2.001
2.999
to
3.001
3.999
to
4.001
4.998
to
5.002
5.998
to
6.002
6.998
to
7.002
7.998
to
8.002
8.998
to
9.002
9.998
to
10.002
9.998
to
10.002
99.9
to
100.1
Page 1 of 1