Navair 17-20VQ-05 Technical Manual page 135

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Fluke 77/AN Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
NOTE;
The TI RANGE HOLD allows a fixed measurement range. Enable the TI RANGE HOLD
as necessary to set the measurement to the TI Range listed in this checklist.
4.1
Direct–Voltage Measurements
TI Range
4.1.3
3.2 V
"
"
"
"
"
"
"
"
4.1.6
–3.2 V
4.1.9
TI RANGE Calibrator
"
320 mV
"
320 mV
"
320 mV
TI Range
4.1.9
32 V
"
320 V
"
1000 V
TI 320 mV Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating Voltage Measurements
TI Range
4.2.2
3.2 V
"
32 V
"
320 V
"
750 V
"
750 V
"
320 V
"
32 V
"
3.2 V
MFG.
NOMINAL
(3)
Calibrator
(V)
1 V x 0
0.000
1 V x .6
0.600
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .3
3.000
(mV)
1 V x .1
100.0
1 V x .2
200.0
1 V x .3
300.0
Calibrator
(V)
10 V x 1
10.00
100 V
100.0
100 V
100
( V)
700
300
Calibrator
(V)
400 Hz–1 V
1.000
400 Hz–10 V
10.00
400 Hz–100 V
100.0
400 Hz–100 V
100
4 kHz–100 V
100
4 kHz–100 V
100.0
4 kHz–10 V
10.00
4 kHz–1 V
1.000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(mV)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.001 to 0.001
0.593
to 0.607
0.989
to 1.011
1.979
to 2.021
2.969
to 3.031
2.969
to 3.031
(mV)
98.9
to 101.1
197.9
to 202.1
296.9
to 303.1
(V)
9.89
to 10.11
98.9
to 101.1
97.7
to 102.3
( V)
600
to 800
200
to 400
(V)
0.958
to 1.042
9.58
to 10.42
95.8
to 104.2
94
to 106
53
to 147
54.5
to 145.5
5.45
to 14.55
0.545
to 1.455