NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Digitec 2110 Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
TI Range
4.3.2
2 k
"
20 k
"
200 k
"
2000 k
TI Range
4.3.2
20 M
118
MFG.
NOMINAL
(3)
Calibrator
( )
100 k
1.000
Calibrator
(k )
1 k
1.000
10 k
10.00
100 k
100.0
1 M
1000
Calibrator
(M )
10 M
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
( )
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
( )
0.997
to
1.003
(k )
0.993
to
1.007
9.93
to
10.07
99.3
to
100.7
992
to
1008
(M )
9.86
to
10.14
Page 2 of 2