Navair 17-20VQ-05 Technical Manual page 113

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8600A Digital Multimeter (Including Options 001 and/or 002)
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
200
TI Range
"
2 k
"
20 k
"
200 k
"
2000 k
TI Range
4.3.2
20 M
Affix a special calibration label stating:
1. The 200 mV V ac range, the V ac frequency response, and the current functions
have not been calibrated.
2. The 200
3. The V ac function has been calibrated up to 50 kHz only.
130
MFG.
NOMINAL
(3)
Calibrator
( )
100
100.00
Calibrator
(k )
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
Calibrator
(M )
resistance range has been calibrated to +(0.6% iv +0.02% of range).
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
( )
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
( )
99.36
to
100.64
(k )
0.9968
to
1.032
9.968
to
10.032
99.68
to
100.32
996.8
to
1003.2
(M )
to
Page 2 of 2