NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Cimron DMM 40 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The TI 0.1 V V dc range, 0.1 V V ac range, 0.1 k range, and all
current functions have not been calibrated.
2. The 1 k range has been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
68
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
10 M
10000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.9980
to
1.0020
9.994
to
10.006
99.94
to
100.06
998.9
to
1001.1
9974
to
10026
Page 2 of 2