Navair 17-20VQ-05 Technical Manual page 70

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Data Tech Corp. 350 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
TI Range Calibrator
4.1.9
1 V
"
100 V
"
1000 V
TI 100 mV Range
4.1.13
10 V LSD
4.1.14
10 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
1 V
"
10 V
"
10 V
"
10 V
"
100 V
"
100 V
"
100 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
V)
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
520
480
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
50 kHz–10 V
10.000
50 kHz–10 V
10.00
4 kHz–10 V
10.00
400 Hz–100 V
100.00
400 Hz–100 V
100.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.001 to
0.001
0.999
to
1.001
1.998
to
2.002
2.998
to
3.002
3.998
to
4.002
4.997
to
5.003
5.997
to
6.003
6.997
to
7.003
7.997
to
8.003
8.996
to
9.004
9.996
to
10.004
9.996
to
10.004
(V)
0.9996 to
1.0004
99.96
to
100.04
99.9
to
100.1
( V)
510
to
530
470
to
490
(V)
0.9987 to
1.0013
9.987
to
10.013
9.987
to
10.013
9.890
to
10.110
9.80
to
10.20
9.80
to
10.20
99.87
to
100.13
99.6
to
100.4
Page 1 of 2
81

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents