CALIBRATION CHECKLIST
TEST INST (S)
Systron Donner 7050 or Fairchild 7050 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
"
15 V
4.1.6
–15 V
4.1.9
1.5 V
"
150 V
"
1000 V
TI 1.5 V Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1.5 k
"
15 k
"
150 k
4.3.2
TI & Calibrator
"
1.5 M
"
15 M
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x 1
10.00
10 V x 1
10.00
1 V x 1
1.000
100 V
100.0
100 V
100
V)
700
300
(k )
range
1.000
range
10.00
range
100.0
(M )
range
1.000
range
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
( V)
(k )
(M )
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.01
to
0.01
0.99
to
1.01
1.99
to
2.01
2.99
to
3.01
3.99
to
4.01
4.98
to
5.02
5.98
to
6.02
6.98
to
7.02
7.98
to
8.02
8.98
to
9.02
9.98
to
10.02
9.98
to
10.02
0.998
to
1.002
99.8
to 100.2
99
to 101
( V)
600
to
800
200
to
400
(k )
0.997
to
1.003
9.97
to
10.03
99.7
to 100.3
(M )
0.989
to
1.011
9.89
to
10.11
Page 1 of 1