NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8800A/AA Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
2 k
"
20 k
"
200 k
"
2000 k
TI Range
4.3.2
20 M
"
200
Affix a special calibration label stating:
1. The TI 200 mV V dc range, and the 2 V V ac range frequency response
have not been calibrated.
2. The TI 200 , 2 k , 20 k
been calibrated to +0.1% of range.
3. The 20, 200 and 1200 V ac Ranges have been calibrated up to 50 kHz only.
100
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.00000
10 k
10.0000
100 k
100.000
1 M
1000.00
Calibrator
(M )
10 M
10.0000
100
100.0
k
and 2000 k resistance ranges have
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL
(4)
(5)
(6)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to
100.200
998.00
to
1002.00
(M )
9.9700
to
10.0300
99.6
to
100.4