CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8800A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
+20 V
4.1.9
–2 V
"
200 V
"
1200 V
TI Range Calibrator
4.1.9
200 mV
TI 200 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
2 V
"
20 V
"
20 V
"
20 V
"
200 V
"
200 V
"
200 V
"
1200 V
"
1200 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x 1
10.0000
1 V x 1
1.00000
100 V
100.000
100 V
100.00
(mV)
1 V x .1
100.000
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.00000
400 Hz–10 V
10.0000
4 kHz–10 V
10.0000
50 kHz–10 V
10.0000
50 kHz–10 V
10.000
4 kHz–10 V
10.000
400 Hz–100 V
100.000
400 Hz–100 V
100.00
4 kHz–10 V
10.00
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
(mV)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.0003 to
0.0003
0.9993
to
1.0007
1.9989
to
2.00011
2.9985
to
3.00015
3.9981
to
4.00019
4.9977
to
5.00023
5.9973
to
6.00027
6.9969
to
7.00031
7.9965
to
8.00035
8.9961
to
9.0039
9.9957
to
10.0043
0.99957 to
1.00043
99.957
to
100.043
99.94
to
100.06
(mV)
99.910
to
100.099
( V)
501
to
503
497
to
499
(V)
0.99690 to
1.0310
9.9690
to
10.0310
9.9690
to
10.0310
9.8440
to
10.1560
9.790
to
10.210
9.960
to
10.040
99.690
to
100.310
99.58
to
100.42
9.85
to
10.15
Page 1 of 2