Navair 17-20VQ-05 Technical Manual page 15

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 34703A Option E01 as used with the Mardan Test Set for FBM*
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
–10 V
"
TI Range
4.1.9
10 mV
"
100 V
TI Range
4.1.9
1 V
"
100 V
"
1000 V
TI 1 volt range
4.1.13
100 V LSD
4.1.14
100 V LSD
*Affix a special calibration label to the TI stating:
"Calibration is performed in accordance with the usage requirement of the Mardan Test Set."
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
Calibrator
(mV)
1000 V
1.000
1 V X .1
100.00
Calibrator
(V)
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
700
300
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(V)
(mV)
(V)
( V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.006
to
0.006
0.991
to
1.009
1.989
to
2.011
2.988
to
3.012
3.986
to
4.014
4.984
to
5.016
5.982
to
6.018
6.980
to
7.020
7.979
to
8.021
8.977
to
9.023
9.975
to
10.025
9.975
to
10.025
(mV)
0.499
to
1.501
99.75
to
100.25
(V)
0.9975
to
1.0025
99.75
to
100.25
99.2
to
100.8
( V)
600
to
800
200
to
400
Page 1 of 1