Navair 17-20VQ-05 Technical Manual page 75

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 3800, 3800A, 3860A, or 429883–01–01 Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
2 k
"
20 k
"
200 k
TI Range
4.3.2
2 M
"
20 M
Affix a special calibration label stating:
"The 200
been calibrated.
NOTE: If the TI is a 3800 or 3800A, the special calibration label should also
state that the V ac function has been calibrated up to 50 kHz only.
86
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.000
10 k
10.00
100 k
100.0
Calibrator
(M )
1 M
1.000
10 M
10.00
range and the direct–current functions of the TI have not
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.997
to
1.003
9.97
to
10.03
99.7
to
100.3
(M )
0.995
to
1.005
9.89
to
10.11
Page 2 of 2

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents