Navair 17-20VQ-05 Technical Manual page 55

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8300A Digital Voltmeter with Options 001 and/or 002
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.2.2
TI Range
"
1000 V
"
1000 V
4.3
Resistance Measurements (8300A–02)
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
4.3.2
10 M
Affix a special calibration label stating:
"The 100 mV V dc range has not been calibrated and all resistance ranges
have been calibrated to +0.2% of range." State also that the V ac frequency
response, and measurement above 50 kHz have not been calibrated; if the TI
has an Option 01 (AC Converter).
64
MFG.
NOMINAL
(3)
Calibrator
(V)
4 kHz–10 V
10.00
50 kHz–10 V
10.00
(k )
range
1.00000
range
10.0000
range
100.000
range
1000.00
(M )
10.000
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
9.59
to
10.41
9.58
to
10.42
Assigned: +0.2%
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to
100.200
998.00
to
1002.00
(M )
9.9800
to
10.0200
Page 2 of 2