Navair 17-20VQ-05 Technical Manual page 123

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
B & K Precision 280 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
4.1.6
–10 V
4.1.9
1000 mV
"
100 V
"
100 V
"
1000 V
4.3
Resistance Measurements
TI Range
4.3.2
100
"
1000
"
1000
Range
"
10 k
"
10 k
"
100 k
"
100 k
*The TI may not indicate this value; however, an overrange indication is considered
acceptable. (Functional Test only)
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x .9
9.00
1 x .9
0.900
10 x 1
10.0
100 V
100*
100 V
100
Calibrator
( )
100
100*
100
100
1 k
1000*
Calibrator
(k )
1 k
1.00
10 k
10.0*
10 k
10.0
100 k
100*
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
ck ( )
( )
ck ( )
ck ( )
(k )
ck ( )
ck ( )
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.06
to
0.06
0.94
to
1.07
1.93
to
2.07
2.92
to
3.08
3.92
to
4.08
4.91
to
5.09
5.91
to
6.09
6.90
to
7.10
7.90
to
8.10
8.89
to
9.11
8.89
to
9.11
0.889
to
0.911
9.8
to
10.2
98.9
to
(101.1)*
93
to
107
( )
97.9
to
(102.1)
88
to
112
979
to
(1021)*
(k )
0.88
to
1.12
9.79
to
(10.21)*
8.8
to
11.2
97.9
to
(102.1)*
Page 1 of 2
143

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents