NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8200, 8200A, or 8200A–BM Digital Voltmeter with Options 001, 002, and/or 003
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (03)
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10 M
Affix a special calibration label stating:
1. The 100
2. All other resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
46
MFG.
NOMINAL
(3)
(k )
1.0000
10.000
100.00
1000.0
(M )
10.000
mV, and 100 mV V dc ranges have not been calibrated.
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.9980
to 1.0020
9.980
to 10.020
99.80
to 100.20
998.0
to 1002.0
(M )
9.980
to 10.020
Page 2 of 2