NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 3465A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range* & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
TI Range* & Calibrator
4.3.2
10 M range
Affix a special calibration label stating:
1. The 10 and 100 mV ranges, and the 100 mV V ac range have not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The DC and AC current functions have not been calibrated.
*In more recent models, range sequence are: 2, 20, 200 etc.
102
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.0000
10.000
100.00
1000.0
(M )
(M )
10.000
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
0.9980
9.980
99.80
998.0
9.980
Page 2 of 2
(7)
(k )
to
1.0020
to
10.020
to
100.20
to
1002.0
(M )
to
10.020