Navair 17-20VQ-05 Technical Manual page 78

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Cimron 7000A, 7200A, 7300A–631, 7400A, or 7500A Digital Multimeter with the 6700A, 6710A
AC Converter, the 6801A or 6802A DC Preamplifier and the 6910A or 6911A Ohm Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
"
+10 V
4.1.6
–10 V
(6801A or 6802A)
TI Range Calibrator
4.1.9
x10
TI Range
"
100 V
"
100 V
"
1000V
(6801A or 6802A:)
TI x100 Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements (6700A or 6701A)
Range
4.2.2
10 V
"
10 V
"
100 V
"
100 V
"
1000 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.0000
10 V x .1
1.0000
10 V x .2
2.0000
10 V x .3
3.0000
10 V x .4
4.0000
10 V x .5
5.0000
10 V x .6
6.0000
10 V x .7
7.0000
10 V x .8
8.0000
10 V x .9
9.0000
10 V x .9
9.0000
mV)
1 x .9
900.00
Calibrator
(V)
10 x 1
10.000
100
100.000
100
100.00
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.0000
4 kHz–10 V
10.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
4 kHz–10 V
10.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
(mV)
(V)
( V)
(V)
ck
ck ( )
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.0001 to
0.0001
0.9998
to
1.0002
1.9997
to
2.0003
2.9996
to
3.0004
3.9995
to
4.0005
4.9994
to
5.0006
5.9993
to
6.0007
6.9992
to
7.0008
7.9991
to
8.0009
8.9990
to
9.0010
8.9990
to
9.0010
(mV)
899.82
to
900.18
(V)
9.998
to
10.002
99.989
to
100.011*
99.98
to
100.02
( V)
501
to
503
497
to
499
(V)
0.9999
to
10.0001
9.9950
to
10.0050
9.999
to
10.001
9.990
to
10.010
9.98
to
10.02
Page 1 of 2