Navair 17-20VQ-05 Technical Manual page 122

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 4600 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
k AUTO
"
"
"
"
"
"
"
"
"
"
Affix a special calibration label stating:
1. The 0.2 V ac & the 1000 V ac range have not been calibrated.
2. The V ac frequency response has been calibrated only at 400 Hz, 4 kHz, and
50 kHz. At 50 kHz, the test instrument is calibrated to: +(0.14% iv + 16 digits).
3. The AC/DC current functions have not been calibrated.
4. The 200
ranges, 2 k through 2 M , have been calibrated to : +(0.18% iv + 1 digit).
142
MFG.
NOMINAL
(3)
Calibrator
(k )
100
0.10000
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
10 M
10000
range has been calibrated to +(0.82% iv + 2 digits) and the remaining
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.09916 to
0.10084
0.9980
to
1.0020
99.800
to
10.020
9.980
to
100.20
998.00
to
1002.0
9979
to
10021
Page 2 of 2