NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 4600 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
k AUTO
"
"
"
"
"
"
"
"
"
"
Affix a special calibration label stating:
1. The 0.2 V ac & the 1000 V ac range have not been calibrated.
2. The V ac frequency response has been calibrated only at 400 Hz, 4 kHz, and
50 kHz. At 50 kHz, the test instrument is calibrated to: +(0.14% iv + 16 digits).
3. The AC/DC current functions have not been calibrated.
4. The 200
ranges, 2 k through 2 M , have been calibrated to : +(0.18% iv + 1 digit).
142
MFG.
NOMINAL
(3)
Calibrator
(k )
100
0.10000
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
10 M
10000
range has been calibrated to +(0.82% iv + 2 digits) and the remaining
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.09916 to
0.10084
0.9980
to
1.0020
99.800
to
10.020
9.980
to
100.20
998.00
to
1002.0
9979
to
10021
Page 2 of 2