CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8000A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
4.1.9
2 V
"
200 V
"
1200 V
TI Range
4.1.9
200 mV
TI 200 mV Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
2 V
"
20 V
"
20 V
"
200 V
"
200 V
"
1200 V
"
1200 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x 1
10.00
1 V x 1
1.000
100 V
100.0
100 V
100
Calibrator
(mV)
1 V x .1
100.0
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.000
400 Hz–10 V
10.00
4 kHz–10 V
10.00
4 kHz–10 V
10.0
400 Hz–100 V
100.0
400 Hz–100 V
100
4 kHz–10 V
10
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
(mV)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.01
to
0.0001
0.99
to
1.01
1.99
to
2.01
2.99
to
3.01
3.99
to
4.01
4.98
to
5.02
5.98
to
6.02
6.98
to
7.02
7.98
to
8.02
8.98
to
9.02
9.98
to
10.02
0.998
to
1.002
99.8
to
100.2
99
to
101
(mV)
99.8
to
100.2
( V)
600
to
800
200
to
400
(V)
0.993
to
1.007
9.93
to
10.07
9.93
to
10.07
9.7
to
10.3
99.3
to
100.7
97
to
103
8
to
12
Page 1 of 2