Navair 17-20VQ-05 Technical Manual page 24

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Dana 4430, 4432, or 4400 Series Digital Voltmeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (with 225 or 230 converter)
TI & Calibrator
4.3
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. All resistance ranges have been calibrated to
2. If the TI is equipped with a model 220 module (converter), the
V ac function has been calibrated up to 50 kHz only.
30
MFG.
NOMINAL
(3)
(k )
range
1.0000
range
10.000
range
100.00
range
1000.0
range
10000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
0.2% of range.
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.9980 to
1.0020
9.980
to
10.020
99.80
to
100.20
998.0
to
1002.0
9980
to 10020
Page 2 of 2

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents