Navair 17-20VQ-05 Technical Manual page 56

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

CALIBRATION CHECKLIST
TEST INST (S)
Tektronix DM 501OPT01, or DM 501OPT02 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
TI Range Calibrator
4.1.9
2 V
"
200 V
"
1000 V
TI 2 V Range
4.1.13
100 V LSD
4.1.14
100 V LSD
4.2
Alternating–Voltage Measurements
Range
4.2.2
2 V
"
20 V
"
20 V
"
20 V
"
200 V
"
200 V
200 V
400 Hz–100 V
"
500 V
"
500 V
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
10 V x 1
10.000
V)
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
700
300
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
50 kHz–10 V
10.000
4 kHz–10 V
10.00
50 kHz–10 V
10.00
100.00
400 Hz–100 V
100.0
50 kHz–10 V
10.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL
(4)
(5)
(6)
(V)
(V)
( V)
(V)
ck ( )
ck ( )
ck ( )
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
(7)
(V)
–0.002 to
0.002
0.997
to
1.003
1.996
to
2.004
2.995
to
3.005
3.994
to
4.004
4.993
to
5.007
5.992
to
6.008
6.991
to
7.009
7.990
to
8.010
8.989
to
9.011
9.988
to
10.012
9.998
to
10.012
(V)
0.9988 to
1.0012
99.88
to
100.12
99.7
to
100.3
( V)
600
to
800
200
to
400
(V)
0.9928 to
1.0072
9.928
to
10.072
9.928
to
10.072
Typical: 9.749 to 10.2
9.928
to
10.072
Typical: 9.75 to 10.25
99.28
to
00.72
99.28
to
100.72
Typical: 9.6 to 10.4
Page 1 of 2
"
65

Advertisement