CALIBRATION CHECKLIST
TEST INST (S)
Hewlett–Packard 3465A Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range*
4.1.3
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
"
10 V
4.1.6
–10 V
4.1.9
1 V
"
100 V
"
1000 V
TI 10 mV Range
4.1.13
1 V LSD
4.1.14
1 V LSD
4.2
Alternating–Voltage Measurements
TI Range*
4.2.2
1 V
"
10 V
"
10 V
"
100 V
"
100 V
"
1000 V
1000 V
*In more recent TI models, range sequences are in terms of: 2, 20, 200, etc.
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.000
10 V x .1
1.000
10 V x .2
2.000
10 V x .3
3.000
10 V x .4
4.000
10 V x .5
5.000
10 V x .6
6.000
10 V x .7
7.000
10 V x .8
8.000
10 V x .9
9.000
10 V x 1
10.000
1 V x 1
1.0000
100 V
100.00
100 V
100.0
V)
502
498
Calibrator
(V)
400 Hz–1 V
1.0000
400 Hz–10 V
10.000
4 kHz–10 V
10.000
4 kHz–10 V
10.00
400 Hz–100 V
100.00
400 Hz–100 V
100.0
4 kHz–10 V
10.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
–0.001
to
0.001
0.999
to
1.001
1.999
to
2.001
2.998
to
3.002
3.998
to
4.002
4.998
to
5.002
5.998
to
6.002
6.998
to
7.002
7.997
to
8.003
8.997
to
9.003
9.997
to
10.003
0.9997
to
1.0003
99.97
to
100.03
99.87
to
100.13
( V)
501
to
503
497
to
499
(V)
0.9980
to
1.0020
9.980
to
10.020
9.945
to
10.055
9.90
to
10.10
99.80
to
100.20
99.3
to
100.7
9.4
to
10.6
Page 1 of 2