Navair 17-20VQ-05 Technical Manual page 65

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Hickok 3300A or 3301 Digital Multimeter (Without Frequency Adapters)
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.2.2
Range
"
1000 V
4.3
Resistance Measurements
TI Range & Calibrator
4.3.2
100
TI Range & Calibrator
"
1 k
"
10 k
"
100 k
TI Range
4.3.2
1 M
"
10 M
"
100 M
Affix a special calibration label on the test instrument stating:
"The AC/DC current function, the V ac frequency response, and
voltage measurements above 50 kHz have not been calibrated."
74
MFG.
NOMINAL
(3)
Calibrator
(V)
400 Hz–100 V
100
( )
100.0
(k )
1.000
10.00
100.0
Calibrator
(M )
1 M
1.000
10 M
10.00
10 M
10.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( )
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
99
to
101
( )
98.7
to
101.3
(k )
0.987
to
1.013
9.87
to
10.13
98.7
to
101.3
(M )
0.996
to
1.004
9.96
to
10.04
9.6
to
10.4
Page 2 of 2