NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Hickok 3300A or 3301 Digital Multimeter (Without Frequency Adapters)
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.2.2
Range
"
1000 V
4.3
Resistance Measurements
TI Range & Calibrator
4.3.2
100
TI Range & Calibrator
"
1 k
"
10 k
"
100 k
TI Range
4.3.2
1 M
"
10 M
"
100 M
Affix a special calibration label on the test instrument stating:
"The AC/DC current function, the V ac frequency response, and
voltage measurements above 50 kHz have not been calibrated."
74
MFG.
NOMINAL
(3)
Calibrator
(V)
400 Hz–100 V
100
( )
100.0
(k )
1.000
10.00
100.0
Calibrator
(M )
1 M
1.000
10 M
10.00
10 M
10.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( )
(k )
(M )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
99
to
101
( )
98.7
to
101.3
(k )
0.987
to
1.013
9.87
to
10.13
98.7
to
101.3
(M )
0.996
to
1.004
9.96
to
10.04
9.6
to
10.4
Page 2 of 2