NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Data Tech Corp. 350 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating the following:
1. The V ac frequency response, the direct–voltage ratio function, the
10, and 100 mV ranges have not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
82
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.00000
10.0000
100.000
1000.00
1000.00
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
0.9980
9.980
99.80
998.0
9980
Page 2 of 2
(7)
(k )
to
1.0020
to
10.020
to
100.20
to
1002.0
to
10020