NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Cimron 7000A, 7200A, 7300A–631, 7400A, or 7500A Digital Multimeter with the 6700A, 6710A
AC Converter, the 6801A or 6802A DC Preamplifier and the 6910A or 6911A Ohm Converter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (6910A or 6911A)
k Ranges
4.3.2
1
"
10
"
100
"
1000
4.3.2
10 M Range
Affix special calibration labels to each of the following Cimron test instruments stating the reasons as
indicated as applicable.
6801A DC Preamplifier: The X100 voltage range has not been calibrated.
6802A DC Preamplifier: The X1000 and X100 voltage ranges have not been calibrated.
7300A, 7300A–631, or 7500A Digital Multimeter: The direct voltage ratio function has
not been calibrated.
6700A or 6710A AC Converter: The 10 kHz V ac response has not been tested.
6710A or 6911A Ohm Converter: The 0.1 k range has not been calibrated.
*The TI may not indicate this value; however, an overrange indication is considered acceptable.
90
MFG.
MODEL
MEASURED VALUES
NOMINAL
FIRST RUN
(3)
(4)
(k )
(k )
1.00000
ck ( )
10.0000
ck ( )
100.000
ck ( )
1000.00
ck ( )
(M )
(M )
10.0000
ck ( )
SER. NO.
OUT
SECOND RUN
CALIBRATION TOLERANCES
OF
TOL
(5)
(6)
0.99950 to
9.9950
99.950
999.50
9.9900
Page 2 of 2
(7)
(k )
1.00050*
to
10.0050*
to
100.050*
to
1000.50*
(M )
to
10.0100*

Advertisement