Navair 17-20VQ-05 Technical Manual page 107

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Cimron 6753 or 6753–1424 Digital Multimeter with the 5775 or 5789 Plug in Unit
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (5789)
TI Range
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating the following information as applicable:
1. The TI direct voltage ratio function has not been calibrated.
2. If the TI consists of a 5789 plug in, the 10 mV and the 100 mV V dc ranges
have not been calibrated; also, all resistace ranges have been calibrated to
+0.2% of range.
3. If the TI consists of a 5775 converter or plug–in, the V ac frequency response,
and measurements above 50 kHz have not been calibrated; also, the 1, 10,
and 100 V V ac ranges have been calibrated to +0.15% of range at 400 Hz.
124
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.00000
10 k
10.0000
100 k
100.000
1 M
1000.00
10 M
10000.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(k )
0.99800 to
1.00200
9.9800
to
10.0200
99.800
to
100.200
998.00
to
1002.00
9980.0
to
10020.0
Page 2 of 2