Navair 17-20VQ-05 Technical Manual

Digital voltmeters and digital multimeters

Advertisement

Quick Links

INSTRUMENT CALIBRATION PROCEDURE
DIGITAL VOLTMETERS
DIGITAL MULTIMETERS
USING THE FLUKE 515A
PORTABLE CALIBRATOR
THIS PUBLICATION SUPERSEDES NAVAIR 17–20VQ–05
DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR
CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR
OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO NAVY
MEASUREMENT SCIENCE DIRECTORATE, NAVAL WARFARE ASSESSMENT STATION, P.O. BOX 5000
CORONA, CA 92878–5000.
DESTRUCTION NOTICE– FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY
METHOD THAT WILL PREVENT DISCLOSURE OF CONTENTS OR "RECONSTRUCTION OF THE
DOCUMENT."
PUBLISHED BY DIRECTION OF COMMANDER NAVAL AIR SYSTEMS COMMAND
NAVAIR 17–20VQ–05
TECHNICAL MANUAL
AND
DATED 1 DECEMBER 1987
15 DECEMBER 1987

Advertisement

Table of Contents
loading

Summary of Contents for Navair 17-20VQ-05

  • Page 1 DIGITAL MULTIMETERS USING THE FLUKE 515A PORTABLE CALIBRATOR THIS PUBLICATION SUPERSEDES NAVAIR 17–20VQ–05 DATED 1 DECEMBER 1987 DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY.
  • Page 2 NAVAIR 17–20VQ–05 LIST OF EFFECTIVE PAGES The purpose of this list is to identify the pages in this document affected by any technical content changes made since the previous release of the document. NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical line in the outer margin of the page.
  • Page 3: Table Of Contents

    NAVAIR 17–20VQ–05 TABLE OF CONTENTS Section Page Introduction and Description ............
  • Page 4: Introduction And Description

    NAVAIR 17–20VQ–05 SECTION 1 INTRODUCTION AND DESCRIPTION 1.1 This procedure describes the calibration of digital voltmeters (DVMs) or digital multimeters (DMMs), and includes instruments having multifunction capability of measuring dc voltage, ac voltage, and resistance. Instruments that can be calibrated using this procedure include, but are not limited to the items listed in Appendix A.
  • Page 5: Equipment Requirements

    If the TI has an electronic counter function, calibration is in two parts: First, calibrate the TI electronic counter function in accordance with NAVAIR 17–20VF–06; then calibrate the multimeter function with this procedure.
  • Page 6: Preliminary Operations

    NAVAIR 17–20VQ–05 SECTION 3 PRELIMINARY OPERATIONS WARNING TESTS INDICATE THAT WHEN USING THE FLUKE 8125A IN EITHER THE BATTERY MODE OR WITH A TWO WIRE V AC INPUT, THERE CAN BE UP TO 180 V AC POTENTIAL BETWEEN THE CASE AND GROUND. ACCORDINGLY, IF THE TI IS A FLUKE 8125A AND A THREE TO TWO...
  • Page 7: Direct Voltage Measurements

    NAVAIR 17–20VQ–05 NOTES Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met before proceeding. Zero test point measurements are to be verified only if an out of tolerance indication occurs at any test point during calibration.
  • Page 8: Alternating-Voltage Measurements

    NAVAIR 17–20VQ–05 4.2 ALTERNATING VOLTAGE MEASUREMENTS 4.2.1 Adjust the TI controls for ac voltage measurements. 4.2.2 Adjust the TI and calibrator controls for TI measurement of the following calibrator voltage and frequency; outputs 400 Hz – 1 V, 400 Hz – 10 V, 400 Hz –100 V, and 4 kHz – 10 V or 50 kHz – 10V. (If the TI has a range above 100 V, it will be tested as 400 Hz –...
  • Page 9 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cubic 50–1122–1, V72–1000–1 Digital Voltmeter as used in Autonetics 37002–315 Static Test Unit NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 10: Navy Calibration Checklist

    NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cubic 0309020 or 0309020–1 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 11 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 2402A with Options 02 and/or 03 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 12 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 2402A with Options 02 and/or 03 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements (03) TI &...
  • Page 13 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron Digital Multimeter, Sanders P/N 3005896G1 as used with 0Q–10/ALM–64* NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 14 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3460B Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 15 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 34703A Option E01 as used with the Mardan Test Set for FBM* NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 16 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3490A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 17 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3490A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements TI & Calibrator (k )
  • Page 18 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Electro Instruments, Inc. 4010 DC Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 19 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–packard 405A, 405AR, 405BR, or 405CR Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 20 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4700 or 4700A Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 21 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4700 or 4700A Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements TI & Calibrator...
  • Page 22 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Non–Linear System 5005 or 5010 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 23 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4430, 4432, or 4400 Series Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 24 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4430, 4432, or 4400 Series Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements (with 225 or 230 converter) TI &...
  • Page 25 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cohu 501B, 501BZ, 502B, or 502BZ Digital Voltmeter with 452, 452A, or 452B AC Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN...
  • Page 26 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5400 to 5403 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 27 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5400 to 5403 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements (07) TI & Calibrator...
  • Page 28 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 6453A, 6453A–1, 6453A–2, or 6453A–3 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 29 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 6453A, 6453A–1, 6453A–2, or 6453A–3 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements TI &...
  • Page 30 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 7000 or 7005A Digital Volt/Multimeter with Options 02, 03, and/or 07 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 31 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 7000 or 7005A Digital Volt/Multimeter with Options 02, 03, and/or 07 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 32 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 7050 or Fairchild 7050 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements...
  • Page 33 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8100A with Options 001, 002, or 8100B Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 34 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8100A with Options 001, 002, or 8100B Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range &...
  • Page 35 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8110A, 8110AOPT001, 8125A, or 8125AOPT001 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 36 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8110A, 8110AOPT001, 8125A, or 8125AOPT001 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI & Calibrator...
  • Page 37 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8200, 8200A, or 8200A–BM Digital Voltmeter with Options 001, 002, and/or 003 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 38 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8200, 8200A, or 8200A–BM Digital Voltmeter with Options 001, 002, and/or 003 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 39 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) AN/USM–216, ME–310, McDonnell 53E04005, or Cubic 93–1015–1 SECTION D Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 40 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) AN/USM–216, ME–310, McDonnell 53E04005, or Cubic 93–1015–1 SECTION D Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI &...
  • Page 41 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8400A with Options 001, 002, 005, and/or 007, 8425A, 8425A/AF, or 8425B Digital Voltmeter with Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED...
  • Page 42 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8400A with Options 001, 002, 005, and/or 007, 8425A, 8425A/AF, or 8425B Digital Voltmeter with Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED...
  • Page 43 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) ME–229/ASM–35 or Electro Instruments, Inc. 851–8014A or 851–8014B Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 44 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) ME–229/ASM–35 or Electro Instruments, Inc. 851–8014A or 851–8014B Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI &...
  • Page 45 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 9000, 9015, or 9025 Digital Volt/Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 46 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 9000, 9015, or 9025 Digital Volt/Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI & Calibrator...
  • Page 47 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 9310, 9320, 9330, 9340, or 9340–5B Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 48 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Systron Donner 9310, 9320, 9330, 9340, or 9340–5B Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements Assigned: +0.2%...
  • Page 49 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8350A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.0000...
  • Page 50 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8350A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements Assigned: +0.2% TI Range & Calibrator...
  • Page 51 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8375A Digital Multimeter with Options 003 and/or 007 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 52 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8375A Digital Multimeter with Options 003 and/or 007 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP 4.2.2 TI Range Calibrator ”...
  • Page 53 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) AN/USM–192 or Simpson 111 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 54 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8300A Digital Voltmeter with Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 55 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8300A Digital Voltmeter with Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP 4.2.2 TI Range Calibrator ”...
  • Page 56 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Tektronix DM 501OPT01, or DM 501OPT02 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 57 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Tektronix DM 501OPT01, or DM 501OPT02 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements Calibrator (k )
  • Page 58 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron DMM 40 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.000...
  • Page 59 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron DMM 40 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator (k )
  • Page 60 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron DMM 50 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.0000...
  • Page 61 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron DMM 50 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI & Calibrator (k )
  • Page 62 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) ME–346/ASM–308, Cimron 9200B, 9300B, 9400B, or 9500B Digital Voltmeter with the 6980B AC Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 63 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) ME–346/ASM–308, Cimron 9200B, 9300B, 9400B, or 9500B Digital Voltmeter with the 6980B AC Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 64 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hickok 3300A or 3301 Digital Multimeter (Without Frequency Adapters) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 65 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hickok 3300A or 3301 Digital Multimeter (Without Frequency Adapters) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP 4.2.2 Range Calibrator ”...
  • Page 66 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Non–Linear System V35, V35A, V35B, V35BR, 3000, 3026, or 3100 Digital Voltmeter with 1201, 1202, 1203, 1204, 125A, 125B, 125C, or 125E AC/DC Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE...
  • Page 67 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Monsanto 200A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.000...
  • Page 68 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3439A or 3440A Digital Voltmeter with the 3441A, 3442A, 3443A, 3444A, 3445A, or 3446A Plug–in Units NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN...
  • Page 69 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3439A or 3440A Digital Voltmeter with the 3441A, 3442A, 3443A, 3444A, 3445A, or 3446A Plug in Units NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN...
  • Page 70 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Data Tech Corp. 350 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 71 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Data Tech Corp. 350 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI & Calibrator (k )
  • Page 72 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hickok 3300 Digital Multimeter (Without Frequency Adapters) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 73 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hickok 3300 Digital Multimeter (Without Frequency Adapters) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP 4.2.2 Range Calibrator ” 1000 V 400 Hz–100 V...
  • Page 74 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 3800, 3800A, 3860A, or 429883–01–01 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 75 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 3800, 3800A, 3860A, or 429883–01–01 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range...
  • Page 76 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5500 as used with Mobil Automatic Test Set LN2C, 5600, 5600OPT20, or 5600–S638 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN...
  • Page 77 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5500 as used with Mobil Automatic Test Set LN2C, 5600, 5600OPT20, or 5600–S638 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN...
  • Page 78 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 7000A, 7200A, 7300A–631, 7400A, or 7500A Digital Multimeter with the 6700A, 6710A AC Converter, the 6801A or 6802A DC Preamplifier and the 6910A or 6911A Ohm Converter NA 17–20VQ–05 SER. NO. PROC. NO.
  • Page 79 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 7000A, 7200A, 7300A–631, 7400A, or 7500A Digital Multimeter with the 6700A, 6710A AC Converter, the 6801A or 6802A DC Preamplifier and the 6910A or 6911A Ohm Converter NA 17–20VQ–05 SER. NO. PROC. NO.
  • Page 80 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 76, 7630, 7650, or 7650–969 Digital Multimeter with the 6770 or 6771 AC Converter, or the 6990 Ohm Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED...
  • Page 81 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 76, 7630, 7650, or 7650–969 Digital Multimeter with the 6770 or 6771 AC Converter, or the 6990 Ohm Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED...
  • Page 82 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8120A Digital Voltmeter including Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 83 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8120A Digital Voltmeter including Options 001 and/or 002 NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements k Ranges:...
  • Page 84 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 9500B–355 Digital Multimeter with the 6980B AC Converter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 85 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dymec DY–2401A, DY–2401AOPTM40, DY–2401B, DY–2410BOPTM40, DY–2401C, and DY–2401COPTM5 Digital Voltmeters NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements (Set TI to 1 Sec sample period)
  • Page 86 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8800A/AA Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 20 V 10 V x 0 0.0000...
  • Page 87 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8800A/AA Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator (k ) (k ) (k ) 4.3.2...
  • Page 88 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3465A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range* Calibrator 4.1.3 10 V 10 V x 0 0.000...
  • Page 89 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 3465A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range* & Calibrator (k )
  • Page 90 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Vidar 520 Integrating Digital Voltmeter* NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 10 V 10 V x 0 0.0000...
  • Page 91 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Vidar 500 Digital Voltmeter Counter* NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 10 V 10 V x 0 0.00...
  • Page 92 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 34740A/34703A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 10 V 10 V x 0 0.000...
  • Page 93 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Data Precision 245 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range* Calibrator 4.1.3 10 V 10 V x 0 0.000...
  • Page 94 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Data Precision 245 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator (k )
  • Page 95 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4300 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.000...
  • Page 96 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4300 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements k Ranges: (k ) (k ) (k ) 4.3.2...
  • Page 97 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4324 Multimeter (with or without Options) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 98 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4324 Multimeter (with or without Options) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements k Ranges: (k )
  • Page 99 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 5306A Multimeter/Counter* NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.0000...
  • Page 100 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 5306A Multimeter/Counter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator (k ) (k ) (k ) 4.3.2...
  • Page 101 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Digitec 2110 Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 V x 0 0.00...
  • Page 102 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Digitec 2110 Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2 100 k 1.000 0.997...
  • Page 103 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) United Systems Corp. (Digitec) 251–3 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator...
  • Page 104 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5000 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.0000...
  • Page 105 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 5000 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements (model 54 ohms Converter) TI Range...
  • Page 106 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 6753 or 6753–1424 Digital Multimeter with the 5775 or 5789 Plug in Unit NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 107 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cimron 6753 or 6753–1424 Digital Multimeter with the 5775 or 5789 Plug in Unit NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 108 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8000A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 V x 0 0.00...
  • Page 109 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8000A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2 TI Range...
  • Page 110 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 970A Probe Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 V–...
  • Page 111 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 970A Probe Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements FUNCTION Calibrator (k ) (k ) (k ) 4.3.2...
  • Page 112 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8600A Digital Multimeter (Including Options 001 and/or 002) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 113 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8600A Digital Multimeter (Including Options 001 and/or 002) NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2...
  • Page 114 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8800A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 V x 0 0.0000...
  • Page 115 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8800A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2 100.00 99.370 100.630...
  • Page 116 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Cubic 242 Instrument System NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +10 V 10 V x 0 0.000...
  • Page 117 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 2000 Danameter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 V x 0 0.00...
  • Page 118 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Electro Instrument 3500A DC Voltmeter/Ratiometer NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 Auto 10 V x 0 0.0000...
  • Page 119 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) United Systems Corporation *2120 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 120 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) United Systems Corporation *2120 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2 100.9...
  • Page 121 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4600 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 DCV AUTO 10 x 0 0.00000...
  • Page 122 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Dana 4600 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator (k ) (k ) (k ) 4.3.2...
  • Page 123 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) B & K Precision 280 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 124 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) B & K Precision 280 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Range Calibrator (M ) (M ) (M ) 4.3.2...
  • Page 125 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Weston 1241 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 x 0 0.00...
  • Page 126 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Hewlett–Packard 2401A, 2401AOPTM40, 2401B, 2401BOPTM40, 2401C, and 2401COPTM5 Digital Voltmeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements (Set TI to 1 Sec sample period)
  • Page 127 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8000A/BU Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator 4.1.3 +20 V 10 V x 0 0.00...
  • Page 128 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8000A/BU Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Resistance Measurements TI Range Calibrator 4.3.2 100.6 TI Range...
  • Page 129 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8500A/AEOPT1/2/5/8 as used with the TS 3846/ASM608 Inertial Measuring Unit Test Set NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements...
  • Page 130 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8500A/AEOPT1/2/5/8 as used with the TS 3846/ASM608 Inertial Measuring Unit Test Set NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES...
  • Page 131 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8010A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP Direct–Voltage Measurements TI Range Calibrator (V dc) (V dc) (V dc) 4.1.3...
  • Page 132 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 8010A Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements TI Range Calibrator (k )
  • Page 133 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Beckman 310 or TECH 310 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Direct–Voltage Measurements TI Range Calibrator 4.1.3...
  • Page 134 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Beckman 310 or TECH 310 Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements 4.3.2...
  • Page 135 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 77/AN Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) NOTE; The TI RANGE HOLD allows a fixed measurement range. Enable the TI RANGE HOLD as necessary to set the measurement to the TI Range listed in this checklist.
  • Page 136 NAVAIR 17–20VQ–05 CALIBRATION CHECKLIST TEST INST (S) Fluke 77/AN Digital Multimeter NA 17–20VQ–05 SER. NO. PROC. NO. MFG. MODEL PROCEDURE MEASURED VALUES FUNCTION TESTED NOMINAL FIRST RUN SECOND RUN CALIBRATION TOLERANCES STEP TOL (6) Resistance Measurements TI Range Calibrator 4.3.2 to 102.2...
  • Page 137 NAVAIR 17–20VQ–05 APPENDIX A REPRESENTATIVE TEST INSTRUMENTS A.1 Test Instrumens that can be calibrated by this procedure include, but are not limited to, the following: Checklist Model Manufacturer Nomenclature Page No. AN/USM–192 Digital Voltmeter ANUSM216 Digital Multimeter DM501 Tektronix ”...
  • Page 138 NAVAIR 17–20VQ–05 Checklist Model Manufacturer Nomenclature Page No. Data Precision Digital Multimeter 251–3 United Systems Corp (Digitec) Digital Voltmeter B & K Precision Digital Multimeter 3000 Non–Linear System Digital Voltmeter 3005896G1 Cimron Digital Multimeter 3026 Non–Linear System Digital Voltmeter Beckman Instruments...
  • Page 139 NAVAIR 17–20VQ–05 Checklist Model Manufacturer Nomenclature Page No. Vidar Digital Voltmeter 53E04005 McDonnell Digital Multimeter 5306A Hewlett–Packard Multimeter/Counter 5400 Dana Digital Voltmeter 5403 ” ” ” 5500 ” ” 5600 ” ” ” 5600OPT20 ” ” ” 5600–S638 ” ”...
  • Page 140 NAVAIR 17–20VQ–05 Checklist Model Manufacturer Nomenclature Page No. 8120A Fluke Digital Multimeter 8120AOPT001 ” ” ” 8120AOPT002 ” ” ” 8125A ” ” 8125AOPT001 ” ” 8200A ” Digital Voltmeter 8200AOPT001 ” ” ” 8200AOPT002 ” ” ” 8200AOPT003 ”...

Table of Contents