NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Data Precision 245 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
TI Range & Calibrator
4.3.2
10 M
Affix a special calibration label stating:
"The current functions have not been calibrated and the 1 k range has been calibrated
to +0.2% of range. The V ac frequency response has not been calibrated."
110
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
(M )
10.000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL
(4)
(5)
(6)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.9980 to
1.0020
9.992
to
10.008
99.92
to
100.08
998.9
to
1001.1
(M )
9.974
to
10.026
Page 2 of 2