Navair 17-20VQ-05 Technical Manual page 94

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Data Precision 245 Digital Multimeter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements
TI Range
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
TI Range & Calibrator
4.3.2
10 M
Affix a special calibration label stating:
"The current functions have not been calibrated and the 1 k range has been calibrated
to +0.2% of range. The V ac frequency response has not been calibrated."
110
MFG.
NOMINAL
(3)
Calibrator
(k )
1 k
1.0000
10 k
10.000
100 k
100.00
1 M
1000.0
(M )
10.000
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL
(4)
(5)
(6)
(k )
(M )
SER. NO.
CALIBRATION TOLERANCES
(7)
(k )
0.9980 to
1.0020
9.992
to
10.008
99.92
to
100.08
998.9
to
1001.1
(M )
9.974
to
10.026
Page 2 of 2