NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Fluke 8375A Digital Multimeter with Options 003 and/or 007
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.2.2
TI Range
"
100 V
"
1000 V
"
1000 V
4.3
Resistance Measurements
TI Range & Calibrator
4.3.2
1000
"
10000
4.3.2
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 0.1 V V dc range has not been calibrated.
2. All resistance ranges have been calibrated to +0.2% of range.
3. The V ac function has been calibrated up to 50 kHz only.
60
MFG.
NOMINAL
(3)
Calibrator
(V)
50 kHz–10 V
10.000
400 Hz–100 V
100.00
4 kHz–10 V
10.00
( )
1000.00
10000.0
(k )
100.000
1000.00
10000.0
MODEL
MEASURED VALUES
FIRST RUN
SECOND RUN
(4)
(5)
(V)
( )
(k )
SER. NO.
OUT
CALIBRATION TOLERANCES
OF
TOL
(7)
(6)
(V)
9.405
to
10.595
99.47
to
100.53
9.59
to
10.41
( )
998.00 to
1002.00
99800
to
10020.0
(k )
99.800 to
100.200
998.00 to
1002.00
9980.0 to
10020.0
Page 2 of 2