Navair 17-20VQ-05 Technical Manual page 31

Digital voltmeters and digital multimeters
Table of Contents

Advertisement

NAVAIR 17–20VQ–05
CALIBRATION CHECKLIST
TEST INST (S)
Systron Donner 7000 or 7005A Digital Volt/Multimeter with Options 02, 03, and/or 07
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.3
Resistance Measurements (03)
TI & Calibrator
4.3.2
1 k
"
10 k
"
100 k
"
1000 k
"
10000 k
Affix a special calibration label stating:
1. The 100 mV V dc range has not been calibrated.
2. All resistance ranges have been calibrated to
3. The V ac function has been calibrated up to 50 kHz only.
38
MFG.
NOMINAL
(3)
(k )
range
1.00000
range
10.0000
range
100.000
range
1000.00
range
10000.0
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
OF
TOL (6)
(4)
(5)
(k )
0.2% of range.
SER. NO.
CALIBRATION TOLERANCES
(7)
Assigned:
0.2%
(k )
0.99800 to
1.00200
9.9800 to
10.0200
99.800
to
100.200
998.00
to
1002.00
9980.0
to 10020.0
Page 2 of 2

Advertisement

Table of Contents
loading

Related Products for Navair 17-20VQ-05

Table of Contents