CALIBRATION CHECKLIST
TEST INST (S)
Dana 2000 Danameter
NA 17–20VQ–05
PROC. NO.
PROCEDURE
FUNCTION TESTED
STEP
NO.
(2)
(1)
4.1
Direct–Voltage Measurements
TI Range
4.1.3
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
"
+20 V
4.1.6
–20 V
4.1.9
2 V
"
100 V
"
1 kV
4.2
Alternating–Voltage Measurements
TI Range
4.2.2
2 V
"
20 V
"
20 V
"
200 V
"
1 kV
"
1 kV
"
200 V
4.3
Resistance Measurements
TI Range
4.3.2
200
20 k
2 M
Affix a special calibration label stating: "The DC current function has not been calibrated."
MFG.
NOMINAL
(3)
Calibrator
(V)
10 V x 0
0.00
10 V x .1
1.00
10 V x .2
2.00
10 V x .3
3.00
10 V x .4
4.00
10 V x .5
5.00
10 V x .6
6.00
10 V x .7
7.00
10 V x .8
8.00
10 V x .9
9.00
10 V x 1
10.00
1 V x 1
1.000
100 V
100.0
100 V
100
Calibrator
(V)
400 Hz–1V
1.000
400 Hz–10 V
10.00
4 kHz–10 V
10.00
4 kHz–10 V
10.0
4 kHz–10 V
10
400 Hz–100 V
100
400 Hz–100 V
100.0
Calibrator
100
100.0
10 k
10.00 k
1 M
1.000 M
MODEL
MEASURED VALUES
OUT
FIRST RUN
SECOND RUN
TOL
(4)
(5)
(6)
(V)
(V)
NAVAIR 17–20VQ–05
SER. NO.
CALIBRATION TOLERANCES
OF
(7)
(V)
–0.01
to
0.01
0.98
to
1.02
1.97
to
2.03
2.97
to
3.03
3.96
to
4.04
4.95
to
5.05
5.94
to
6.06
6.94
to
7.06
7.93
to
8.07
8.92
to
9.08
9.91
to 10.09
0.994 to 1.006
99.1to 100.9
99to 101
(V)
0.972to 1.028
9.72to 10.28
9.72to 10.28
9.4to 10.6
8 to 12
96to 104
97.2to 102.8
96.7to 103.3
9.88to 10.12 k
0.969to 1.031 M
to
Page 1 of 1
(136 blank) 135