Chapter 2: General settings
Header: Select ON to print the signal and recording information on header of the
recorder paper. Select OFF to print only signal on the paper (without any information).
Recorder Setting Menu
Select Rec Setting from the Main Menu to access the below menu:
The following items can be set through this menu:
Rec Time: To set recording duration for different leads in Auto mode. It ranges from 3 to
12 seconds.
Periodic Rec. Interval: To set time interval in periodic recording. Available options are 5-
60 min and Off. If you select Off, periodic recording will be stopped.
Periodic Rec. Repetition: Press to set repetition of recording. Available options are 1-20
and ∞.
Sensitivity: To set amplitude of ECG waveform in record & display screen. Available
options are 2.5, 5, 10 and 20 (mm/mV).
Paper Speed: Press to set the recording speed. Available options are 6.25, 12.5, 25 and 50
(mm/sec).
Figure 2-3 Recorder setting menu
16