7 SCPI Programming Reference
CALibrate:SAVE
Saves the calibration constants in non-volatile memory. Do this at the end of the calibration to avoid
losing changes.
Parameter
(none)
Store cal constants in non-volatile memory: CAL:SAVE
If calibration mode is exited by programming CALibration:STATe OFF without first saving the new
l
constants, the previous calibration constants are restored.
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands.
The optional parameter is the password. This is required if the existing password is not 0 (zero). If the
password is not entered or is incorrect, an error is generated and the calibration mode remains
disabled.
Parameter
0|OFF|1|ON , default OFF
<password> a numeric value up to 15 digits
Disable calibration mode and unsecure calibration: CAL:STAT OFF, 0
Enable calibration using a numeric password: CAL:STAT ON, 1234
If calibration mode is exited by programming CALibration:STATe OFF without first saving the new
l
constants, the previous calibration constants are restored.
CALibrate:VOLTage[:LEVel] <value>, (@<channel>)
Calibrates the output voltage. The value entered selects the range that is being calibrated.
Parameter
The maximum voltage of the range.
Calibrates the voltage of the 50 V range: CAL:VOLT 51, (@1)
CALibrate:VOLTage:CMRR (@<channel>)
Calibrates the voltage common mode rejection ratio.
270
Typical Return
(none)
Typical Return
0 or 1
(none)
Typical Return
(none)