Download Print this page

Keysight Technologies N9030B Installation Note

Pxa signal analyzer, option cr3 second if out/crp arbitrary if out/alv log video out
Hide thumbs Also See for N9030B:

Advertisement

Quick Links

Keysight N9030B PXA Signal Analyzer
Option CR3 Second IF Out
Option CRP Arbitrary IF Out
Option ALV Log Video Out
Installation Note

Advertisement

loading

Summary of Contents for Keysight Technologies N9030B

  • Page 1 Keysight N9030B PXA Signal Analyzer Option CR3 Second IF Out Option CRP Arbitrary IF Out Option ALV Log Video Out Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Option CR3, or CRP, or ALV Installation Kit Option CR3, or CRP, or ALV Installation Kit Products Affected: N9030B PXA Signal Analyzer Serial Numbers: Options To Be Performed By: (X) Keysight Service Center (X) Personnel Qualified by Keysight (X) Customer Estimated Installation Time: 0.5 Hours...
  • Page 4 This kit contains all of the instructions required to install Option CR3, CRP, or ALV on a Keysight N9030B PXA signal analyzer. The rear panel AUX IF OUT port is present on all PXA signal analyzers, however each of the possible output signal choices require a license.
  • Page 5: Installation Procedure Over Usb

    — Test cable, sma or 3.5 mm connectors — Adapters as needed to connect the cable between the spectrum analyzer and sma AUX IF OUT port on the rear panel of the N9030B. For Option ALV — Signal source capable of providing pulse output above 200 MHz —...
  • Page 6: Verify The Installation

    Option CR3, or CRP, or ALV Installation Kit Figure 1 USB Storage Device Configuration Menu 7. The signal analyzer will automatically consume the License File. (This may take a few minutes) When the License File is consumed the Keysight License Manager will display a “Successful License Installation”...
  • Page 7 Option CR3, or CRP, or ALV Installation Kit Functional Check: Within one year of the initial calibration date of the instrument this option is fully calibrated with no further adjustment or verification testing. To ensure that this newly installed option has been enabled correctly the following functional check is recommended.
  • Page 8 Option CR3, or CRP, or ALV Installation Kit Span: 1 MHz Bandwidth: 100 kHz The spectrum analyzer should display a −33 dBm ± 5 dB signal Option ALV, Fast Log Video 1. Preset X-Series analyzer. Assure mode is Spectrum Analyzer. 2.
  • Page 9 For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...
  • Page 10 This information is subject to change without notice. © Keysight Technologies 2016 Edition 1, February 2016 N9030-90079 www.keysight.com...