Keysight Technologies N524 A Series Installation Note
Keysight Technologies N524 A Series Installation Note

Keysight Technologies N524 A Series Installation Note

2-port/4-port , pna-x upgrade kit
Table of Contents

Advertisement

Quick Links

Keysight 2-Port/4-Port
N524xA to N524xB and
N5264A to N5264B Model
PNA-X Upgrade Kit
This manual provides documentation for the following models: N524xA and
N5264A
To Upgrade 2-Port/4-Port "A" models to 2-Port/4-Port "B" models
N5241AU-690 (For N5241B), N5242AU-690 (For N5242B), N5244AU-690 (For
N5244B), N5245AU-690 (For N5245B), N5247AU-690 (For N5247B), N5249AU-690
(For N5249B), and N5264AU-690 (For N5264B)
Upgrade Kit Numbers: N5242-60117, N5245-60115, N5247-60114, N5264-60102
Installation Note

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies N524 A Series

  • Page 1 Keysight 2-Port/4-Port N524xA to N524xB and N5264A to N5264B Model PNA-X Upgrade Kit This manual provides documentation for the following models: N524xA and N5264A To Upgrade 2-Port/4-Port "A" models to 2-Port/4-Port "B" models N5241AU-690 (For N5241B), N5242AU-690 (For N5242B), N5244AU-690 (For N5244B), N5245AU-690 (For N5245B), N5247AU-690 (For N5247B), N5249AU-690 (For N5249B), and N5264AU-690 (For N5264B) Upgrade Kit Numbers: N5242-60117, N5245-60115, N5247-60114, N5264-60102...
  • Page 2 WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS documentation. No additional DOCUMENT THAT CONFLICT WITH government requirements © Keysight Technologies, Inc. THESE TERMS, THE WARRANTY beyond those set forth in the 2007-2019 TERMS IN THE SEPARATE EULA shall apply, except to the AGREEMENT WILL CONTROL.
  • Page 5 Keysight Licensed Functionality Upgrade Kit Upgrade Kit Numbers: N5242-60117, N5245-60116, N5247-60114, N5264-60102 Installation Note Description of this Upgrade This upgrade converts your N52xxA microwave network analyzer or N5264A microwave receiver to an equivalent N52xxB or N5264B. This is done by removing the existing front panel assembly, outer covers, and installing a new front panel assembly, SSDA, and outer covers.
  • Page 6 Description of this Upgrade Getting Assistance from Keysight Getting Assistance from Keysight Installing this upgrade kit requires special skills and experience. If you think you may not be qualified to do the work, or need advice, contact Keysight. If You Have Problems With the Upgrade Kit Contents Keysight stands behind the quality of the upgrade kit contents.
  • Page 7 Description of this Upgrade Getting Prepared Getting Prepared License Key File Redemption IMPORTANT! As part of the upgrade, the service center needs to request a the customer’s Keysight License files from Keysight. Refer to “License Key File Redemption” on page IMPORTANT! Also, you will need to contact the Keysight factory support node to request multiple license key files.
  • Page 8 Description of this Upgrade Getting Prepared 2. In the Keysight License Manager welcome window, click You can add a new certificate. 3. In the "Keysight Software Manager: Add Certificate" window that opens, in the "Enter your certificate information" section, refer to your Software License Certificate that you received with your upgrade kit and enter the following: —...
  • Page 9 Obtain Multiple License Key Files All of the licenses in your instrument need to be converted to a new format for your "B" model instrument. Contact Keysight Technologies factory support early in your upgrade process to receive your licenses in a timely manner. See “Getting Assistance from Keysight”...
  • Page 10 Description of this Upgrade Protecting Your Workspace from Electrostatic Discharge Downloading the Online PNA Service Guide To view the Service Guide for your PNA model number, use the following steps: 1. Go to www.keysight.com. 2. In the Search box, enter the model number of the analyzer (Ex: N5242B) and click Search.
  • Page 11 Description of this Upgrade Tools Required for the Installation Tools Required for the Installation Description T-10 TORX driver - set to 9 in-lbs (1.02 N.m) T-20 TORX driver - set to 21 in-lbs (2.38 N.m) 5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m) 5/16-in (8 mm) nutsetter or open end torque wrench - set to 21 in-lbs (2.38 N.m) 9 mm nutsetter or open end torque wrench - set to 21 in-lbs) (2.38 N.m) 1/4-in (6 mm) open end wrench...
  • Page 12: Lower Front Panel Overlay, 2-Port (N5241/2/9B Option

    Items Included in the Upgrade Kit Items Included in the Upgrade Kit Check the contents of your kit against the following list. If any part is missing or damaged, contact Keysight Technologies. Refer to “Description of this Upgrade” on page This section applies to multiple models and upgrade kits.
  • Page 13: Table Of Contents

    Description of this Upgrade Items Included in the Upgrade Kit Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B), N5242AU-690 (N5242B), and N5249AU-690 (N5249B) Description Part Number Lower front panel overlay, 2-port (N5241/2/9B Option 224 with Option 029) N5242-80027 Lower front panel overlay, 4-port (N5241/2/9B Option 419/423 with Option 029) N5242-80028 Lower front panel overlay, 2-port (N5241/2/9B Option 217)
  • Page 14 Description of this Upgrade Items Included in the Upgrade Kit Table 1 Contents of Upgrade Kit N5242-60117 for N5241AU-690 (N5241B), N5242AU-690 (N5242B), and N5249AU-690 (N5249B) Description Part Number N5221-00007 Test set, subpanel (4-port) - N5241/2/9B only N5242-60009 Wire assy, LED PCA - N5241/2/9B only Synthesizer board (With tabs for s/n 5200 and above) N5240-60074...
  • Page 15 Description of this Upgrade Items Included in the Upgrade Kit Table 2 Contents of Upgrade Kit N5245-60116 for N5244AU-690 (N5244B) and N5245AU-690 (N5245B) Description Part Number Installation Note (this document) N5242-90023 Software Entitlement Certificate 5964-5145 Front handle 5023-3074 Rear foot 5041-7903 Bottom foot 5041-7906...
  • Page 16 Description of this Upgrade Items Included in the Upgrade Kit Table 2 Contents of Upgrade Kit N5245-60116 for N5244AU-690 (N5244B) and N5245AU-690 (N5245B) Description Part Number Lower Dress Panel, 4-Port (ALL PNA models). Goes on, before lower front panel N5240-00009 overlay.
  • Page 17 Description of this Upgrade Items Included in the Upgrade Kit Table 3 Contents of Upgrade Kit N5247-60114 for N5247AU-690 (N5247B)) Description Part Number Installation Note (this document) N5242-90023 Software Entitlement Certificate 5964-5145 Front frame – (assembly) N5240-60100 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619 Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel overlay frame to 0515-1946...
  • Page 18 Description of this Upgrade Items Included in the Upgrade Kit Table 3 Contents of Upgrade Kit N5247-60114 for N5247AU-690 (N5247B)) Description Part Number Outer cover N5245-00035 Guard, jumper cables, side—2-Port & 4-Port N5242-00048 Guard, jumper cables, center—4-Port N5242-00049 Guard, jumper cables, center—4-Port N5247-00024 Front impact cover N9040-40007...
  • Page 19: N5242

    Description of this Upgrade Items Included in the Upgrade Kit Table 4 Contents of Upgrade Kit N5264-60102 for N5264AU-690 (N5264B) Description Part Number Installation Note (this document) N5242-90023 Software Entitlement Certificate 5964-5145 Front frame – (assembly) N5240-60071 Machine screw, M4.0 x 25 pan head (To attach rear feet to chassis.) 0515-1619 Machine screw, M3.0 x 10 mm flat head (To attach lower dress panel overlay frame to 0515-1946...
  • Page 20: N5242

    Description of this Upgrade Initial Instrument Inspection Initial Instrument Inspection STOP! This repair must be done at a service center! Refer to “Getting Assistance from Keysight” on page This section and all steps must be completed before proceeding to the following sections or customer data may be lost or instrument repair delayed! The purpose of this section is to verify where the instrument should begin the...
  • Page 21: N5242

    Description of this Upgrade Initial Instrument Inspection Table 6 Overview of Assembly Requirements for B Model Upgrade Assembly Part Number Serial Number for PNAs that start using Can be these boards (PNAs with these serial ordered prefixes do not need to be upgraded) using the part number, CPU, version 7...
  • Page 22: N5242

    Description of this Upgrade Initial Instrument Inspection Verifying the Board Assemblies The following board assemblies need to be verified and installed during the upgrade: 1. Press Help > About NA... 2. In the About Network Analyzer window that opens, make a note of the following: —...
  • Page 23: N5242

    Description of this Upgrade Initial Instrument Inspection 6. If your instrument has the minimum criteria mentioned above or you have received the required parts proceed to “Step 1. Copy All Unique Files to an External Disk Drive” on page Overview of the Installation Procedure Windows XP OS only: Before installing the new SSDA with Windows 7 OS—...
  • Page 24: N5242

    Description of this Upgrade Initial Instrument Inspection All instruments: “Step 1. Copy All Unique Files to an External Disk Drive” on page 24 Windows 7 Operating System (Windows 7 OS) only: “Step 2. Update Windows 7 OS Instruments with A.10.40 or above” on page 25 “Step 3.
  • Page 25: N5242

    Description of this Upgrade Initial Instrument Inspection “Step 18. Install the New Handler I/O Cable” on page 44 “Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit” on page 45 “Step 20. Reboot the Instrument” on page 46 “Step 21.
  • Page 26: Step 1. Copy All Unique Files To An External Disk Drive

    Description of this Upgrade Step 1. Copy All Unique Files to an External Disk Drive Step 1. Copy All Unique Files to an External Disk Drive The hard disk drive (HDD/SSD) in your analyzer will be replaced as part of this upgrade.
  • Page 27: Step 2. Update Windows 7 Os Instruments With A.10.40 Or Above

    Description of this Upgrade Step 2. Update Windows 7 OS Instruments with A.10.40 or above Step 2. Update Windows 7 OS Instruments with A.10.40 or above Skip this step, if your PNA is already running firmware A.10.40 or above. 1. If the PNA is running, click File > Exit 2.
  • Page 28: Step 3. Verify Windows 7 Os With New A.10.40.00 Firmware Or Greater

    Description of this Upgrade Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater Step 3. Verify Windows 7 OS with New A.10.40.00 Firmware or Greater Skip this step, if your PNA is already running firmware A.10.40 or above. 1.
  • Page 29: Step 4. Remove The Outer Covers

    Description of this Upgrade Step 4. Remove the outer covers Step 4. Remove the outer covers Remove the Outer Cover 1. Turn OFF the analyzer and disconnect the power cord. ➀ 2. With a T-20 TORX driver, remove the strap handles (item ) by loosening ➁...
  • Page 30: Step 5. Remove The Power Supply Cover

    Description of this Upgrade Step 5. Remove the Power Supply Cover Step 5. Remove the Power Supply Cover 1. On the rear panel, remove the screws (x7) and the power supply cover. Save the 0515-0372 (x7) screws for reuse. 2. If you do not have a new Hi-Pot label, then you need to peel off the old Hi-Pot sticker and save for reuse.
  • Page 31: Step 7. Install The Service Center's N5242-60133 Solid-State Disk Drive Assembly (Ssda) With A.10.40 Or Above-(Windows Xp Os Only)

    Description of this Upgrade Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only) Step 7. Install the Service Center’s N5242-60133 Solid-State Disk Drive Assembly (SSDA) with A.10.40 or above—(Windows XP OS Only) IMPORTANT! If your instrument already had a Windows 7 operating system and already had a Version 7 CPU prior to this upgrade, skip to...
  • Page 32: Step 8. Install The *New* Power Supply Cover

    Description of this Upgrade Step 8. Install the *New* Power Supply Cover Step 8. Install the *New* Power Supply Cover Refer to Figure 4 Figure 5 for this step. 1. Install the new power supply cover by reusing the 0515-0372 (x7) screws. 2.
  • Page 33: Step 9. Power Up Your Instrument With The *New* Version 7 Cpu And Allow Windows 7 To Stabilize

    Description of this Upgrade Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7 to Stabilize Step 9. Power Up Your Instrument with the *New* Version 7 CPU and Allow Windows 7 to Stabilize IMPORTANT! If your instrument already had a Windows 7 operating system and already had a Version 7 CPU prior to this upgrade, skip to “Step 10.
  • Page 34: Step 10. Remove The Front Panel Assembly

    Description of this Upgrade Step 10. Remove the Front Panel Assembly Step 10. Remove the Front Panel Assembly ➀ 1. With a 5/16-inch wrench, remove all semi-rigid jumpers (item ) from the front panel. s ➁ 2. With a T-10 TORX driver, remove the eight 0515-0244 screws (item from the sides of the frame.
  • Page 35: Step 11. Inspect And If Required, Replace Front Subpanel

    Description of this Upgrade Step 11. Inspect and If Required, Replace Front Subpanel Step 11. Inspect and If Required, Replace Front Subpanel IMPORTANT! If your front panel has the screw holes shown in Figure 7 on page 33, skip this step. STOP! If you have a N5244/5A and it is necessary to replace the front subpanel, you will also need to rotate the LED boards and replace the bias cables.
  • Page 36: Step 12. If Required, Rotate The Led Boards And Replace The Led Bias Cables (N5244/5A Only)

    Description of this Upgrade Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only) Step 12. If Required, Rotate the LED Boards and Replace the LED Bias Cables (N5244/5A Only) IMPORTANT! If the PNA is a N5244A/45A and it was necessary to replace the front subpanel, you will also need to rotate the LED boards 180 degrees and replace the bias cables.
  • Page 37: Step 13. Inspect Or Modify Midplane Board (W1312-63095 To A W1312-63376)

    Description of this Upgrade Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376) Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376) If your instrument already has the correct W1312-63376 midplane board, skip to “Step 14. Install the Dress Panel on the New Front Panel Assembly” on page STOP!!! Please read all content before proceeding with this upgrade procedure.
  • Page 38: N5240

    Description of this Upgrade Step 13. Inspect or Modify Midplane Board (W1312-63095 to a W1312-63376) Verify that the sticker has the "W1312-63376" part number. Refer to Figure If not, and you have read and understand the Caution on page 35, then refer to Appendix A: “Midplane Board Upgrade (W1312-63095 to a W1312-63376),”...
  • Page 39: Step 14. Install The Dress Panel On The New Front Panel Assembly

    Description of this Upgrade Step 14. Install the Dress Panel on the New Front Panel Assembly Step 14. Install the Dress Panel on the New Front Panel Assembly Install the new lower dress panel onto the new front panel using the 10 flat head screws (0515-1946) provided.
  • Page 40: Step 15. Install The New Overlays On The New Front Panel Assembly

    Description of this Upgrade Step 15. Install the New Overlays on the New Front Panel Assembly Step 15. Install the New Overlays on the New Front Panel Assembly IMPORTANT! To avoid possible damage to the lower front panel overlay (label), do not attempt to attach the lower front panel label until “Step 16.
  • Page 41: Step 16. Install The New Front Panel Assembly And The Lower Overlay

    Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay Step 16. Install the New Front Panel Assembly and the Lower Overlay Before installing the front panel assembly onto the analyzer, place the analyzer right side up and lift and support the front of the analyzer chassis. ➁...
  • Page 42: N5240

    Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay 3. Hand start the twelve screws (item 0515-2044) in the sides of the frame. Use the new screws from the kit and those retained earlier. Do not torque. Refer to Figure 10 on page 39 Figure...
  • Page 43: N5242

    Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay 7. Install each of the overlays as follows (refer to Figure 12): — Remove the protective backing to expose the adhesive. — Carefully align the overlay to its location and gently position it. —...
  • Page 44: N5242

    Description of this Upgrade Step 16. Install the New Front Panel Assembly and the Lower Overlay 8. If equipped with front panel jumpers, reinstall the semi-rigid jumpers (Figure 13) on the front panel using a 5/16 inch torque wrench set to 8 in-lbs (N5247B only) and 10 in-lbs (N5241/2/4/5/9B).
  • Page 45: Step 17. Install New Version Spam V, Synth, If Multiplexer, And Test Set Mother Boards (If Necessary)

    Description of this Upgrade Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards (If Necessary) Step 17. Install New version SPAM V, Synth, IF Multiplexer, and Test Set Mother boards (If Necessary) A) Install the new SPAM V Boards If your PNA already has the SPAM V boards, skip to next step.
  • Page 46: Step 18. Install The New Handler I/O Cable

    Description of this Upgrade Step 18. Install the New Handler I/O Cable Step 18. Install the New Handler I/O Cable Refer to your instrument Service Guide’s section "Remove and Replace the A19 Test Set Motherboard" on www.keysight.com. 1. Remove the A19 TSMB and be careful to mark all of the cables before removal (i.e., If your TSMB needs to be replaced ("N524063043"), discard this older board.) 2.
  • Page 47: Step 19. Install The *New* N5242-60135 Solid-State Disk Drive Assembly (Ssda) From The Upgrade Kit

    Description of this Upgrade Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit Figure 15 Fold the Handler I/O Cable (8121-2919 and N5242-60005) Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit 1.
  • Page 48: Step 20. Reboot The Instrument

    Else, if the unit still has initialization errors which look like problems that cannot be resolved, contact the Keysight Technologies, because the model and serial number need to be updated via the Keysight network. Refer to “Getting Assistance from Keysight”...
  • Page 49: Step 21. Reinstall The Outer Covers

    Description of this Upgrade Step 21. Reinstall the Outer Covers Step 21. Reinstall the Outer Covers Reinstall the Outer Cover 1. Slide the cover onto the frame. ③ 2. Using a T-20 TORX driver, install the four rear panel feet (item ) by ➃...
  • Page 50: Step 22. Install Option N52Xxau-690

    Successful License Installation is displayed on the bottom of the window. If your PNA does not automatically reboot, contact Keysight Technologies. “Getting Assistance from Keysight” on page 4. If the license file installed and the PNA rebooted as expected, remove your USB memory device.
  • Page 51: N5245

    Description of this Upgrade Step 22. Install Option N52xxAU-690 Option Verification Procedure When you power up your instrument there will be many errors, because your instrument does not have any licenses installed. This is as expected and will be corrected in later in this procedure. Once the analyzer has restarted and the Network Analyzer program is again running: 1.
  • Page 52: Step 23. Changing Your Pna's Computer Name From An A Model To A B Model

    Description of this Upgrade Step 23. Changing your PNA’s Computer Name from an A model to a B model Step 23. Changing your PNA’s Computer Name from an A model to a B model If the PNA does not have a B model after the upgrade, it is necessary to run the ChangeHostname program to change the PNA from an A model to a B model.
  • Page 53: Step 24. Install And Verify Multiple License Files

    Description of this Upgrade Step 24. Install and Verify Multiple License Files Step 24. Install and Verify Multiple License Files After completing this process all of your license files should be loaded into your instrument. Procedure Requirements — The analyzer must be powered up and operating to perform this procedure. —...
  • Page 54: Step 25. Verify The Pna Analyzer Program Is Running With The Correct Options

    Description of this Upgrade Step 25. Verify the PNA Analyzer Program is Running with the Correct Options 7. Disconnect the USB flash drive from the PNA. Step 25. Verify the PNA Analyzer Program is Running with the Correct Options If the options have been enabled, skip to “Step 26.
  • Page 55: Step 26. Copy The Unique Files Back To The Analyzer's Ssd

    Description of this Upgrade Step 26. Copy the Unique Files Back to the Analyzer’s SSD Step 26. Copy the Unique Files Back to the Analyzer’s SSD The files that were previously saved onto a USB flash memory drive must now be installed onto the new SSD. —...
  • Page 56: Step 27. Perform Post-Upgrade Adjustments And Calibration

    Description of this Upgrade Step 27. Perform Post-Upgrade Adjustments and Calibration Step 27. Perform Post-Upgrade Adjustments and Calibration Adjustments The following adjustments must be made due to the hardware changes of the analyzer. — EE default adjustment (Only, if the synthesizer board is changed) —...
  • Page 57: Step 28. Prepare The Pna For The User

    Description of this Upgrade Step 28. Prepare the PNA for the User Step 28. Prepare the PNA for the User 1. If necessary, reinstall front jumper cables. 2. Install the cable guards, pushing them over the front jumper cables until the cushioning material touches the front panel of the PNA.
  • Page 58: N5247

    Description of this Upgrade Step 28. Prepare the PNA for the User Installation Note N5242-90023...
  • Page 59: N5240

    Keysight Licensed Funcitionality Upgrade Kit N52xxA Microwave Network Analyzers Installation Note Midplane Board Upgrade (W1312-63095 to a W1312-63376) STOP!!! Please read all content before proceeding with this upgrade procedure. This is a complex procedure that requires Keysight support training, before beginning any repairs! If you have not been properly trained by Keysight support personnel, attempting to do this procedure could result in damage to the midplane board and or the instrument! See also...
  • Page 60: N5240

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Table 1 Parts List for Midplane Board Upgrade Kit Modification Part number Description 0699-1429 Resistor U403—383 ohm (attach to pin 5) 8150-3431 30 gauge wire (0.5 inch) 5184-4489 Label-Brady blank label 0470-0954 Loctite 495 a.
  • Page 61: N5242

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure Procedure 1. To remove your midplane board, refer to the Service Guide for your instrument that can be downloaded from www.keysight.com. 2. Cut pin 5 on U403, so that half of the lead is left near on the IC. Refer to Figure Figure 17 Step 1 Cut pin 5 half way up as shown—Top Side View...
  • Page 62: N5247

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure 3. Clean pad and remove cut off portion of U403–pin 5 from the pad. Refer to Figure Figure 18 Cleaning the pad for pin 5—Top Side View N5242-90023 Installation Guide...
  • Page 63: N5240

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure 4. Add 383 ohm resistor as shown. Refer to Figure Figure 19 Add 383 ohm resistor N5242-90023 Installation Guide...
  • Page 64: W1312

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure 5. Add a 1/2 inch 30 gauge wire. Refer to Figure Figure 20 Add a 1/2 inch 30 gauge wire—Top Side View N5242-90023 Installation Guide...
  • Page 65: N5242

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure 6. Add glue as shown to hold the wire in place. Refer to Figure 21 on page Figure 21 Add glue as shown to hold the wire in place—Top Side View N5242-90023 Installation Guide...
  • Page 66: N5242

    Midplane Board Upgrade (W1312-63095 to a W1312-63376) Procedure 7. Add the W1312-63376 sticker. Verify that the sticker does not interfere with the PCA screw holes. Refer to Figure Figure 22 Adding sticker to new board number W1312-63376—Top Side View 8. Re-install the midplane board per the Service Guide for your instrument that can be downloaded from www.keysight.com.
  • Page 67: N5247

    Keysight Licensed Funcitionality Upgrade Kit N52xxA Microwave Network Analyzers Installation Note Manually Setting the PNA Display to 24-Bit This section is only required if, after installing the new SSD that comes with this upgrade kit (refer to “Step 19. Install the *NEW* N5242-60135 Solid-State Disk Drive Assembly (SSDA) from the Upgrade Kit”...