3 Using the Source Functions
CURR:LIM 1,(@1)
To set the output voltage of all outputs to 10 V:
VOLT 10.02,(@1:4)
To enable only output 1:
OUTP ON,(@1)
To enable output 1 and output 3:
OUTP ON,(@1,3)
To measure the output voltage and current of output 1:
MEAS:VOLT? (@1)
MEAS:CURR? (@1)
Configuring an Output Turn-On/Turn-Off Sequence
Turn-on and turn-off delays control the turn-on and turn-off timing of the outputs in relation to each
other.
You can also synchronize Output On/Off delays across multiple mainframes. Refer to
Output Couple Controls
Step 1 – Set the output voltage and current of the output channels:
Refer to steps 1 and 2 under
all outputs that will be sequenced.
Step 2 – Configure the turn-on turn-off delays:
Press the Settings key twice to access the Output On/Off Delays window. Enter the On Delays and Off
Delays for all outputs that will participate in the output on/off delay sequence. Values can range from
0 to 1023 seconds.
82
for further information.
Controlling the Outputs
and set the output voltage and current values of