Repairing the Detector

Nano Flow Cell - Replacing or Cleaning

When required
Tools required
Parts required
For details refer to the technical that comes with the nano-flow cell kit.
N O T E
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows
N O T E
at the quartz block.
1 Disconnect the capillaries from the capillary holder and
remove the flow cell.
122
If parts are contaminated or leaky.
Screwdriver POZI 1 PT3
Two 1/4 inch wrenches for capillary connections
For parts identification refer to
500 nl).
2 Unscrew the cell body from the holder.
1100 Series DAD and MWD Reference Manual
"Nano Flow Cells"
on page 192 (80 nl and