Agilent Technologies 1100 Series Reference Manual page 111

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

8
Insert a new gasket [6, 7] and the window assembly [1]
into the cell body. Assure, that the gasket is on the bottom.
1
7
Next steps:
Reconnect the capillaries, see
Perform a leak test.
Insert the flow cell.
Replace the front cover
Perform a wavelength calibration, see
Test"
on page 86 to check the correct positioning of the flow cell.
1100 Series DAD and MWD Reference Manual
6
1
Do not mix the gasket #6
and # 7 (different hole
diameter)
"Exchanging a Flow Cell"
"Wavelength Recalibration"
9
Using a 4-mm hex key, tighten the window screw hand
tight plus a quarter turn.
on page 105.
on page 84 or holmium test, see
Repairing the Detector
"Holmium Oxide
4
111