Agilent Technologies 1100 Series Reference Manual page 106

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector
2
Press the release button and open the flow cell door.
4
Remove the flow cell by pressing the flow cell holder.
106
3
Disconnect the cell inlet and the cell outlet capillary and
the waste capillary from the capillary holder.
Note:
The label attached to the flow cell provides
information on part number, path length and
maximum pressure.
If you want to replace flow cell parts, see
the Standard or Semi-Micro Flow Cell"
or
"Repairing the Standard or Semi-Micro Flow
Cell"
on page 108.
1100 Series DAD and MWD Reference Manual
"Repairing
on page 108