2 Pull the filter motor off the shaft. Keep the spring in a safe
place.
4 While counter-holding the filter lever, press the filter
motor onto the filter shaft until it clicks into the final
position.
1100 Series DAD and MWD Reference Manual
3 Before installing the filter motor, press the filter lever in
from the other side.
Next steps:
•
Check that the filter lever is moveable.
•
Reinstall the optical unit, see
(G1315A/G1365A)"
on page 171 or
Optical Unit (G1315B/G1365B)"
•
Reconnect the filter motor cable to connector J18
(G1315/65A) or J19 (G1315/65B) on the detector main
board.
•
Reinstall the front cover, top cover and top foam section,
see
"Replacing the Foam and the Top Cover"
page 175.
•
Replace the detector into the stack.
•
Reconnect the power cable and turn on detector.
Repairing the Detector
"Installing the Optical Unit
"Removing the
on page 156.
on
4
163