Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Pull the filter motor off the shaft. Keep the spring in a safe
place.
4 While counter-holding the filter lever, press the filter
motor onto the filter shaft until it clicks into the final
position.
1100 Series DAD and MWD Reference Manual
3 Before installing the filter motor, press the filter lever in
from the other side.
Next steps:
Check that the filter lever is moveable.
Reinstall the optical unit, see
(G1315A/G1365A)"
on page 171 or
Optical Unit (G1315B/G1365B)"
Reconnect the filter motor cable to connector J18
(G1315/65A) or J19 (G1315/65B) on the detector main
board.
Reinstall the front cover, top cover and top foam section,
see
"Replacing the Foam and the Top Cover"
page 175.
Replace the detector into the stack.
Reconnect the power cable and turn on detector.
Repairing the Detector
"Installing the Optical Unit
"Removing the
on page 156.
on
4
163