4
Repairing the Detector

Entering the Serial Number using the Control Module

1 Connect the control module to the detector. Turn on the detector.
2 In the control module, press System (F5), then Records (F4). Using the
3 Press FW Update (F5). Now, press the m key. This will display a box which
4 Press Enter. This will display the box labeled Serial#.
5 Alphabetic characters and numbers are created using the up and down
For firmware revisions below A.02.00 it is important never to press Done if the Serial# box
N O T E
is blank. In this case, the module can no longer be recognized by either the control module
or the Agilent ChemStation. The main board must then be replaced.
6 Turn the detector off, then on again. The Records screen should display the
7 If a Agilent ChemStation is also connected, restart the Agilent ChemStation
146
up/down arrows, make sure that the detector is highlighted.
says Update Enter Serial#.
arrows. In the box labeled Serial#, enter the 10-character serial number for
the detector. When the 10-character serial number is entered, press Enter
to highlight the complete serial number. Then, press Done (F6).
correct serial number for this module.
now as well.