Optional Interface Boards

The Agilent 1100 Series modules have one optional board slot that allows
addition of an interface board to the modules.
Table 33
Description

BCD Board

Fuse 250 mA (four are on the board)
LAN Communication Interface Board
BCD Board
The BCD board provides a BCD output for the bottle number of the
Agilent 1100 Series autosampler and four external contacts. The external
contact closure contacts are relay contacts. The maximum settings are: 30 V
(AC/DC); 250 mA (fused).
Figure 64
1100 Series DAD and MWD Reference Manual

Optional Interface Boards

Board
identification
Processor
BCD register
interface
Block Diagram BCD Board
Line driver
+
External
contacts
Introduction to the Detector
Part Number
G1351-68701
2110-0004
G1369A or G1369-60001
12
BCD
connector
250 mA
External contact
4 x
connector
6
253