Repairing the Detector
13 Fix the new heat exchanger to the clamp unit.
118
Next steps:
Reconnect the capillaries, see
Cell"
on page 105.
Perform a leak test.
Insert the flow cell.
Replace the front cover.
Perform a wavelength calibration, see
Recalibration"
on page 84 or holmium test, see
"Holmium Oxide Test"
positioning of the flow cell.
1100 Series DAD and MWD Reference Manual
"Exchanging a Flow
"Wavelength
on page 86 to check the correct