The Main Power Supply Assembly - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

6
Introduction to the Detector

The Main Power Supply Assembly

The main power supply comprises a closed assembly (no onsite repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from 100 – 120 or
220 – 240 volts AC ± 10 % and needs no manual setting.
l i n e
v o l t a g e
Figure 67
To disconnect the instrument from line, unplug the power cord. The power supply
WA R N I N G
still uses some power, even if the power switch on the front panel is turned off.
266
M P S
p r i m a r y
c i r c u i t s
s e n s e
Main Power Supply (MPS) Blockdiagram
+ 3 6 V
s u p p l y
+ 2 4 V
s u p p l y
+ 5 V
s w i t c h i n g
r e g u l a t o r
l i n e a r v o l t a g e
r e g u l a t o r s
1100 Series DAD and MWD Reference Manual
+ 3 6 V ( n o t u s e d )
+ 2 4 V / 4 . 5 A
+ 5 V / 3 A
+ 1 5 V / . 3 A
- 1 5 V / . 3 A