Advertisement

If you want to replace the gasket only, continue with
step 8.
3 If the washers fall out of the window assembly, they must be inserted in the correct order with the Teflon
any leaks from the flow cell window.
Follow the procedure
Orientation of Flow Cell Parts
1 window screw
contains items 2, 3, 4, 5
and 6
2 seal ring
3 quartz window
4 compression washer
5 spring washers
4 window holder
6 window screw
7 flow cell
8 inlet capillary
9 outlet capillary
1100 Series DAD and MWD Reference Manual
"Repairing the Standard or Semi-Micro Flow Cell"
6
5
2 Use a tooth pick to remove the quartz window from the
window assembly.
4
2
3
2
8
9
Repairing the Detector
®
ring to prevent
on page 108 for reassembling.
7
4
1
113