Repairing the Detector
Exchanging the Source Lens (Achromat) Assembly
When required
Tools required
Parts required
If the filter motor was removed, the filter lever should not be reused. Always use a new
N O T E
filter lever to assure correct fit on the filter motor shaft.
Preparations for this procedure:
•
Turn off the lamp.
•
Switch off detector and disconnect the power cable.
•
Remove detector from stack and place it on the working
bench.
•
Remove the front cover, top cover and top foam section,
see
"Removing the Top Cover and Foam"
•
Remove the optical unit, see
(G1315A/G1365A)"
Optical Unit (G1315B/G1365B)"
158
The source lens (achromat) assembly might be cleaned or replaced when the
transmission of light is reduced and a lamp exchange or new spectro
windows do not improve the light throughput
Screwdriver POZI 1 PT3
hexagonal key 1.5 mm
A pointed pair of pliers
Source lens (achromat) assembly G1315-65201(includes filter lever
G1315-45001)
Spectro window 79880-28111
Filter lever G1315-45001
on page 140.
"Removing the Optical Unit
on page 154 or
"Removing the
on page 156.
1 Place the optical unit on your bench.
1100 Series DAD and MWD Reference Manual