Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

No accessible hardware fuse is needed because the main power supply is safe
against any short circuits or overload conditions on the output lines. When
overload conditions occur, the power supply turns off all output voltages.
Turning the line power off and on again resets the power supply to normal
operation if the cause of the overload condition has been removed.
An over-temperature sensor in the main power supply is used to turn off
output voltages if the temperature exceeds the acceptable limit (for example, if
the cooling fan of the instrument fails). To reset the main power supply to
normal operating conditions, turn the instrument off, wait until it is
approximately at ambient temperature and turn the instrument on again.
The following table gives the specifications of the main power supply.
Table 48
Maximum power
Line Input
Output 1
Output 2
Output 3
Output 4
Output 5
1100 Series DAD and MWD Reference Manual
Main Power Supply Specifications
300 VA / 130 W
100 – 120 or 220 – 240 volts AC
± 10 %, line frequency of 50/60 Hz
+ 24 V / 4.5 A (maximum)
+ 36 V / 2.5 A (maximum)
+ 5 V / 3 A
+ 15 V / 0.3 A
- 15 V / 0.3 A
Introduction to the Detector
Continuous output
Wide ranging
total power consumption of + 24 V
and + 36 V must not exceed 107 W.
6
267