Removing The Optical Unit (G1315B/G1365B) - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Detector

Removing the Optical Unit (G1315B/G1365B)

When required
Tools required
Parts required
With the introduction of the G1315B/G1365B DAD/MWD the appearance of the z-plane
N O T E
(front metal plane) has been changed (no holes any more) and an isolation seal between
the optical unit and the z-plane. Therefore, some of the figures may still show the old
appearance of the G1315A/G1365A. For G1315A/G1365A detectors refer to
Optical Unit (G1315A/G1365A)"
Preparations for this procedure:
Turn off the lamp.
Switch off detector, and disconnect power cable.
Remove detector from stack and place it on working
bench.
Remove front cover, top cover and top foam section, see
"Removing the Top Cover and Foam"
156
For all repairs inside the optical unit
Screwdriver POZI 1 PT3
Optical unit G1315-69002 (exchange assembly) or
Individual parts depending on the following procedures
on page 154.
1 .Loosen the screw on the left side of the z-plane and the
on page 140.
two screes near the hydraulic connection.
1100 Series DAD and MWD Reference Manual