Exchanging A Flow Cell - Agilent Technologies 1100 Series Reference Manual

Diode array and multiple wavelength detectors
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Exchanging a Flow Cell

When required
Tools required
Parts required
Preparations for this procedure:
Turn the lamp(s) off.
1100 Series DAD and MWD Reference Manual
If an application needs a different type of flow cell or the flow cell needs
repair.
Two 1/4 inch wrenches for capillary connections
Standard flow cell, 10 mm, 13 µl, 120 bar, G1315-60012
Semi-micro flow cell, 6 mm, 5 µl, 120 bar, G1315-60011
High pressure flow cell, 6 mm, 1.7 µl, 400 bar, G1315-60015
Nano flow cell, refer to
"Nano Flow Cell - Replacing or Cleaning"
1
Press the release buttons and remove the front cover to
have access to the flow cell area.
Repairing the Detector
on page 122
4
105