4
Repairing the Detector
7 Replace the cell seal assembly onto the quartz body.
Always use a new seal assembly to exclude damage
during disassembling.
9 Insert the flow cell capillaries and tighten them
fingertight. Use the wrench and torque adapter as
described on
page 126
124
and tighten the fittings alternately.
8 Slide the quartz body completely into the cell body to the
front stop (use for example a toothpick).
10 Reassemble the flow cell body to the holder.
1100 Series DAD and MWD Reference Manual